Fluke have issued a voluntary recall notice for their 37X range of clamp meters, manufactured between 1st September 2010 to 1st October 2012. The affected models and serial numbers are:

  • Fluke 373 – 14270001 to 21950000
  • Fluke 374 – 14270001 to 16379999
  • Fluke 375 – 14270001 to 16079999
  • Fluke 376 – 14270001 to 15909999

Note serial numbers ending in “”R”” are not affected and do not need to be returned

The Issue

The printed circuit assembly may not be properly fastened to the test lead input jack. This may result in inaccurate voltage readings, including a low or no voltage reading on a circuit energised with a hazardous voltage, presenting a shock, electrocution or thermal burn hazard.

Anyone with a clamp that is affected is advised to stop using it and return it to Fluke for immediate replacement. Fluke will handle the collection of your affected clamp and the despatch of your new one.

What to do next

You can check if your clamp is affected and organise the return of this directly with Fluke, please visit their website for full details.